CQM Resistance - The CQM resistance can be entered directly be pressing on the resistance indicator
button. The maximum CQM resistance is 500 ohms.
Change by Ohms - These buttons allow the user to adjust the CQM resistance up or down by a set
number of ohms. The maximum step size is 500 ohms.
Change by Percent - These buttons allow the user to adjust the CQM resistance up or down by a
percentage of the current resistance. Maximum percent change is 500%. The minimum resistance
change is 1 ohm. That is, if the current resistance is 1 ohm and the % change is 1%, pressing the Up /
Down buttons will still change the resistance by 1 ohm.
Zero Ohms - This button will set the CQM resistance to zero ohms.
Open Circuit - This button will toggle the CQM resistance to an open circuit.
The RF Leakage screen allows the user to take RF leakage measurements. The measurement mode can
be configured using the keys around the top and right side of the screen.
The buttons on this screen are described below in a clockwise sequence starting with the back arrow.
MEASURE RF LEAKAGE