Leakage 2 - This leakage test, specified by the IEC as Earth Reference Leakage Type BF (Load Between
Electrodes), is for testing the leakage to earth ground of a Ground Referenced Output type BF
electrosurgical generator from the active output. This test complies with IEC 601.2.2, sec. 19.101a, test 1,
fig. 102, and sec. 19.102, adopted by ANSI/AAMI HF18-2001. The purpose of this test is to verify that the
RF leakage of the Device Under Test (DUT) meets or exceeds the IEC specification.
Leakage 3 - This leakage test, specified by the IEC as Earth Reference Leakage Type BF (Load from
Active Electrode to Earth) is for testing the leakage to earth ground of a Ground Referenced Output type
BF electrosurgical generator from the active output. This test complies with IEC 601.2.2, sec. 19.101a,
test 2, fig. 102, and sec. 19.102, adopted by ANSI/AAMI HF18-2001. The purpose of this test is to verify
that the RF leakage of the Device Under Test (DUT) meets or exceeds the IEC specification.
Test Setup - This key shows a picture of the connections for the current mode.