Display Parameters - This key allows the user to select the number of measurements, or zones, that are
shown on the screen. 6 Screen configurations are available, 5 Display Screens which have 1, 2, 3, 4 and
5 display zones respectively, and a Measurement List Screen which shows all available measurements.
Each Display Zone can be customized to show the desired parameter from the following options:
Show Graph - This key allows the user to display a graph of the latest RF measurement. Graphs can be
saved and loaded from internal memory or external USB storage.
Averaging - This key allows the user to select the FAST, MEDIUM, or SLOW averaging mode. Fast
averaging will provide a quick response to incoming signals. Slow averaging will provide a more stable
display, but will be slower to respond to small changes in the RF input signal. The value for the Averaging
settings can be configured in the System Setup screen.
Input Range - Allows the user to select the range for the RF input. The range can be selected as
100 mV, 1000 mV, or Auto range.
Mode Parameter Abbreviation Description
Pulsed Mode Only
Time - Pulse On
T
on
This is the duration that the pulsed waveform is on
(see Diagram 1)
Time - Pulse Off
T
off
This is the duration that the pulsed waveform is off
(see Diagram 1)
Time - Total Cycle
T
cyc
This is the total cycle time of the pulsed waveform
(i.e. T
on
+ T
off
) (see Diagram 1)
% Duty Cycle %Duty
This is the ratio of the pulse on time (T
on
)
versus cycle time (T
cyc
) (see Diagram 1)
mV Pulse
mV
cyc
This represents the RMS mV over one pulsed cycle
(see Diagram 1)
mA Pulse
mA
cyc
This represents the RMS mA over one pulsed cycle
(see Diagram 1)
Watts Pulse
W
cyc
This represents the RMS Watts over one pulsed cycle
(see Diagram 1)
Ton
Toff
Tcyc
mV
mA
Watts
CF
mVcyc
mAcyc
Wcyc
Diagram 2 (Pulsed mode only)