PC1 Connection PC2 ConnectionI
Sequence
Protection class I devices
: The protective conductor test must already
have been passed as a prerequisite for the insulation resistance test.
Ð Select the test: ↑↓ keys.
Ð Select the protection class and the type of connection:
PC1 / PC2 / FIX. key.
Ð
Connect the DUT to the test socket, and connect the probe if necessary.
Note
All switches at the device under test must be set to the on posi-
tion during measurement of insulation resistance, including
temperature controlled switches and temperature regulators as
well. Measurement must be performed in all program steps for
devices equipped with program controllers.
Ð Start the test: Press the STARTSTOP key.
Attention!
!
Testing is conducted with up to 500 V. Current limiting is uti-
lized (I < 10 mA), but if the terminals (L and N) are touched,
electrical shock may occur which could result in conse-
quential accidents.
Note: Open-circuit voltage is always greater than nominal voltage.
Ð PC2 connection: Contact exposed conductive parts with the
probe during measurement.
Ð All measured values are displayed.
Ð End the test: Press the STARTSTOP key.
Ð Read the measured value and compare it with the table of
permissible limit values.
Examples of Minimum Permissible Limit Values for Insulation Resistance
Permanent connection
Test
Standard
Test Volt a ge
R
ISO
PC I PC II PC III Heat
IEC 62353
(VDE 0751-1)
500 V
2MΩ 7MΩ
70 MΩ 70 MΩ
VDE0701-0702
1MΩ 2MΩ 0.25 MΩ 0.3 MΩ