Operation Manual
22 GMC-I Messtechnik GmbH
I
T
Touch Current – Testing for Absence of Voltage
Sequence DIR / DIF / ALT
Ð Select the test: ↑↓ keys.
Ð Connect the DUT to the test socket, or connect the probe.
Ð Select type of test current: DIR / DIF / ALT key.
Ð Select mains polarity reversal: L/N / N/L key.
Ð Start the test: Press the START⏐STOP key.
Ð Measured values are displayed.
Ð End the test: Press the START⏐STOP key.
Ð Read the measured value and compare it with the table of
permissible limit values.
Examples of Maximum Permissible Limit Values for
Touch Current in mA
Procedure for DL – 2-probe Measurement
This measurement is performed with 2 probes. The measuring
section is electrically isolated from the mains power supply of the
instrument. Input resistance is 1 kΩ.
Ð Select test: key ↑↓
Ð Connect probe 1 (e. g. the 25 m cable drum) to sockets 1-2
and connect the probe tip with the reference measuring point.
Ð Select test current type: key DL
Ð Scan the test point with probe 2 (socket connectors 3-4).
Ð Start test: press key START⏐STOP.
Ð Measured values are displayed.
Ð Quit test: Press key START⏐STOP.
Ð Read off measured value and compare it with the table of per-
missible limit values.
Test Standard
Protec-
tion
Class
Direct /
Differential Current
Measurement
Alternative Measurement
IEC 62353
(VDE 0751-1)
PC2 0.1 mA 0.5 mA
VDE 0701-702 PC2 0.5 mA