Operation Manual
GMC-I Messtechnik GmbH 27
I
AP
Leakage Current from the Application Part (alternative patient leakage current, mains at applied part)
Mains Polarity Parameter
Polarity can be reversed for measurements during which the
mains are connected to the test socket.
Note
Can only be used for types BF and CF applied parts.
Ð Select the test: ↑↓ keys.
Ð Connect the device under test to the test socket and the ap-
plied parts to the patient ports. The test probe has to be con-
nected but without applying electrical contact (potential-free).
Ð Select type of test current: DIR / ALT key.
Ð Select mains polarity reversal: L/N / N/L key.
Ð Select applied parts A through K: → key.
Ð Start the test: Press the START⏐STOP key.
Ð Measured values are displayed.
Ð End the test: Press the START⏐STOP key.
Ð Read the measured value and compare it with the table of
permissible limit values.
Examples of Maximum Permissible Limit Values for Leakage
Current in mA
Test Standard AP
Direct Measurement
(mains at AP)
Alternative Measurement
(alternative patient leakage
current)
IEC 62353
(VDE 0751-1)
BF 5 mA 5 mA
CF 0.05 mA 0.05 mA
IEC 60601
BF 5 mA —
CF 0.05 mA —
IEC 60601 3rd ed.
Total Patient
Leakage Current
BF 5 mA —
CF 0.1 mA —