Operation Manual

GMC-I Messtechnik GmbH 33
Technical Data
Electrical Safety
Fuses 2 x FF (UR) 500 V/16 A AC;
6,3 mm x 32 mm;
(Article number 3-578-215-01)
50 kA breaking capacity at 500 V AC
Safety class Disconnection from mains per SC II
Nominal voltage 230 V
Test voltage 2.2 kV AC or 3.3 kV DC
Measuring category 300 V CAT II
Fouling factor 2
Safety Shutdown With following differential current at DUT
during:
– Function test
10 mA~ / < 25 ms
Touch current meas.
direct current meas.
10 mA~ / < 25 ms
Residual current meas.
20 mA~ / < 25 ms
– Protective conductor
direct current meas.
10 mA~ / < 25 ms
Residual current meas.
20 mA~ / < 25 ms
with following probe current during:
– Touch current meas. 10 mA~ / < 5 ms
– Protective conductor
resistance measurement
300 mA~ / < 1ms
Electromagnetic Compatibility, EMC
Interference Emission
EN 61326-1:2006 class B
Interference Immunity
EN 61326-1:2006
Mechanical Design
Display monochrome backlit dot matrix display,
128 x 128 pixels
Dimensions (W x D x H) 325 x 250 x 90 mm
Weight approx. 2 kg
Protection Housing: IP 40, connections: IP 20
per DIN VDE 0470 part 1/EN 60529
Table Excerpt Regarding Significance of the
IP Code
Data Interface
USB Slave
IP XY
(1
st
digit X)
Protection against pene-
tration of solid particles
IP XY
(2
nd
digit Y)
Protection against
penetration by water
0 Not protected 0 Not protected
1 50.0 mm dia. 1 vertically falling drops
2 12.5 mm dia. 2
vertically falling drops with
enclosure tilted 15°
3 2.5 mm dia. 3 spraying water
4 1.0 mm dia.
4 Splashing water