Safety Warnings
• Power Consumers with High Inrush Current (> 16 A) – Function Test
(e.g. fluorescent tubes, halogen lamps, headlights etc.):
Observe the following instructions in order to prevent
excessive contact loads.
Attention!
!
Starting the Function Test
For reasons of safety, the device under test must be
switched off before the function test is started. This
precaution prevents inadvertent start-up of a device under
test which may represent a hazard during operation, e.g. a
circular saw or a disc grinder.
Ending the Function Test
After completion of the function test, devices under test
must be turned off with their own switch – especially
devices with relatively high inductivity.
The measuring instrument may not be used:
• If it demonstrates visible damage
• With damaged connector cables, measuring cables or
patient ports
• If it no longer functions properly
In such cases, the instrument must be removed from operation
and secured against unintentional use.
Meanings of Symbols on the Instrument
300 V CAT II
Maximum permissible voltage and measuring category
between connections 1 through 4, the test socket and
ground
I System with maximum 16 A nominal current
Warning regarding dangerous electrical voltage
Warning concerning a point of danger
(attention: observe documentation!)
The device may not be disposed of with the trash.
Further information regarding the WEEE mark can be
accessed on the Internet at
www.gossenmetrawatt.com by entering the search
term WEEE.
Opening of Equipment / Repair
The equipment may be opened only by authorized service per-
sonnel to ensure the safe and correct operation of the equipment
and to keep the warranty valid.
Even original spare parts may be installed only by authorized ser-
vice personnel.
In case the equipment was opened by unauthorized personnel,
no warranty regarding personal safety, measurement accuracy,
conformity with applicable safety measures or any consequential
damage is granted by the manufacturer.
!