8.8 Measuring Time to Trip for RCDs, Type PRCD – tA
Schematic Diagram 
Definition
According to DIN VDE 0100, part 610:2004, substantiation must 
be provided that RCCBs are tripped at nominal residual current 
within 400 ms (1000 ms for selective RCCBs).
PRCD: Portable residual current device (only those whose 
protective conductor is not deactivated)
Application
The PRCD under test is plugged into the test socket at the test 
instrument. The PRCD’s phase conductor must be contacted 
with test probe P1 in order to trip the PRCD. This is usually the 
left-hand socket.
Test Sequence
➭ Set the rotary switch to the t
A
 position.
➭ Plug the PRCD into the test socket at the test instrument and 
connect the test probe to P1.
➭ Start the test: press the START/STOP key. 
➭ Acknowledge the warning which indicates that line 
voltage will be connected to the test socket. 
Execute the following steps when prompted to do so:
Note
Please note that test probe P1 is in continuous contact 
with the phase conductor from the point in time at which 
the PRCD is plugged in until it trips. Premature discon-
nection of the test probe may result in erroneous mea-
sured values.
➭ After each reconnection to the mains, and as soon as the first 
test is started, a mains connection test is executed.
➭ If the probe test has revealed that probe P1 was not con-
nected: connect probe P1 as described above.
➭ Switch the PRCD on within 5 seconds or activate it (e.g. reset 
button on PRCD), so that the PRCD can be activated when 
energized with line voltage.
➭ Contact neutral conductor L at the PRCD with test probe P1 
(ascertain by trial and error if necessary).
➭ The test is automatically ended and time to trip is displayed 
after the PRCD is tripped.
Note
If the PRCD was not switched on, or if N was contacted 
instead of L at the PRCD with test probe P1, an error 
message is displayed and “Error” appears at the display.
➭ The save symbol appears and prompts you to save the 
measured values to an ID number.
➭ Press the ESC key in order to discard the measured 
values stored to buffer memory and acknowledge by 
pressing the key shown at the right. 
Switch 
Setting
Meas. 
Variant
Measuring Function
Single measurements, rotary switch level: green
Measurements at DUTs with line voltage
t
A
t
B
PRCD time to trip for 30 mA PRCD
U
LN
Line voltage at the test socket
t
A