User Manual
46 GMC-I Messtechnik GmbH
12 Characteristic Values
1
Specified values are only valid for the display at the test instrument. 
Data transmitted via the USB port may deviate from these values.
2
Known as equivalent leakage current or equivalent patient leakage current from 
previous standards
3
  Protective conductor current, touch current, device leakage current, patient 
leakage current
4
Protective conductor current, touch current, device leakage current
5
Only with SECUTEST BASE10 and SECUTEST XTRA with feature G01
Key: rdg. = reading (measured value), d = digit(s)
Test Times, Automated Sequence
Test time > 2 s, exception: device protective conductor resistance 
RPE: > 7 s. Test times are not checked or calibrated, but rather 
determined on the basis of processor cycle times.
Emergency Shutdown During Leakage Current Measurement
As of 10 mA of differential current (can also be set to 30 mA), 
automatic shutdown ensues within 100 ms.
Func-
tion
Measured Quan-
tity
Measuring Range 
/
Nominal Range of 
Use
Reso-
lution
Nominal
Voltage
U
N
Open-Cir-
cuit Volt-
age
U
0
Nom. 
Current
I
N
Short-
Circuit
Current
I
K
Int. 
Resis-
tance
R
I
Ref. 
Resis-
tance
R
REF
Measuring 
Uncertainty 
1
Intrinsic Error 
1
Overload 
Capacity
Value Time
Tests, 62638 (DIN VDE 0701-0702) / IEC 62353 (VDE 0751)
Protective 
conductor 
resistance 
R
PE
000 999 m 1m
—
< 24 V 
AC or DC
—
>200 mA
AC or DC
> 10 A AC 
5
——
(15% rdg.+ 10 d)
>10d
(10% rdg.+ 10 d)
>10d
264 V
250 mA
16 A 
5
Cont.
1.00 ... 9.99  10 m
10.0 30.0 
100 m
Insulation 
Resistance
RINS
10  999 k 1k
50  500 
VDC
1.0 • U
N
 
1.5 • U
N
>1mA <2mA — —
(5% rdg.+ 4 d)
>10d
(2.5 % rdg.+2 d)
> 10 d
264 V Cont.
1.00  9.99 M 10 k
10.0  99.9 M 100 k
20 M:
(10% rdg.+ 8 d)
20 M:
(5% rdg.+4 d)
100  300 M 1M
Leakage current, 
alternative 
measurement 
2
IPE, IB, IG, IA
0.0 ... 99 A1A
—
50 ... 
250 V~
– 20/+10%
—<1.5mA
> 150 k
1k
10 
(5% rdg.+ 4 d)
> 15 mA: 
(10% rdg.+ 8 d)
(2 % 
rdg.
+2 D)> 10 d
> 15 mA: 
(5% rdg.+ 4 d)
264 V Cont.
100 ... 999 A1A
1.00 to 9.99 mA 10 A
10.0 to 30.0 mA 100 A
Leakage current, 
d
irect 
measurement 
3
IPE, IB, IG, IA, IP
Only Ip: 0.0 ... 
99.9 A
100 nA
————
1k
10 
—
(5% rdg.+ 4 d) 
> 10 d
(2.5 % rdg.+2 d)
> 10 d
264 V Cont.
0.0 ... 99 A1A
100 ... 999 A1A
1.00 to 9.99 mA 10 A
10.0 to 30.0 mA 100 A
Leakage current, 
d
ifferential 
current meas. 
4
IPE, IB, IG
0 ... 99 A1A
————
1k
10 
—
(5% rdg.+ 4 d)
> 10 d
(2.5 % rdg.+2 d)
>10d
264 V Cont.
100 ... 999 A1A
1.00 to 9.99 mA 10 A
10.0 to 30.0 mA 100 A
Function Test
Line voltage 
U
L–N
100.0  240.0 V~ 0.1V— — ———— —
(2% rdg.+2 d)
264 V Cont.
Load current I
L
0  16.00 A
RMS
10mA— — ———— —
(2% rdg.+2 d)
16 A Cont.
Active power P
0  3700 W 1 W — — — — — — —
(5 % rdg.+10 d)
>20d
264 V Cont.
20 A 10 min
Apparent power S
0  4000 VA 1 VA Calculated value, U
L–N
 • I
V
(5 % rdg.+10 d)
>20d
Power factor PF
with sinusoidal 
waveform: cos
0.00  1.00 0.01 Calculated value, P / S, display > 10 W (10% rdg.+5 d)
U-
P
ROBE
Probe voltage
(phase search)
, and 
0.0  99.9 V 100 mV
—— — — —
(2% rdg.+2 d)
264 V Cont.
100 to 300 V 1 V
t
A
PRCD
Time to trip at 
30 mA
0.1 ... 999 ms 0.1 ms — — — — — —
5 ms