User manual

GMC-I Gossen-Metrawatt GmbH 15
4 Adjusting Output Values
4.1 Output Voltage
4.1.1 Uout – Presently Measured Voltage Value
The left-hand digital display [4] <Volt> indicates the measured
value Uout for voltage present at the output terminals.
During sensing mode operation, the displayed value
corresponds to voltage acquired from the sensing leads
at the consuming device.
The 3½ place measured value display is refreshed approxi-
mately 3 times per second. Overflow is indicated for measured
values of greater than 19.99 at 20 V devices.
If output voltage is superimposed with an alternating voltage,
the arithmetic mean value is displayed.
Refer to chapter 1.4.3, Electrical Data, regarding measuring
range, measuring resolution and measuring accuracy.
4.1.2 Ulim – Limit Value for Uset
Function
Defines the voltage adjustment setting range <Uset> [8]
Prevents inadvertent, excessively high voltage setpoint values
(Uset) during manual operation.
If Ulim is set to 0 V (full left), the voltage adjusting
potentiometer <Uset> is disabled (e.g. for adjustment of output
voltage by means of the analog interface).
Setting Ulim
Ulim is adjusted with the left-hand trimming potentiometer [7]
<Ulim>.
Use a size 3 screwdriver only in order to avoid damaging the
potentiometer.
Deactivate the output: <OUTPUT> ”OFF”.
First set Uset to its maximum value (turn potentiometer [8] <Uset>
clockwise as far as it will go).
Press the key [9] <DISPLAY Uset/Iset>.
The left-hand display [4] <Volt> is switched from the presently
measured voltage value Uout to the manually selected voltage
setpoint Uset.
Hold this key depressed.
At the same time, turn <Ulim> [7] with the screwdriver until <Volt>
appears at the display [4] as the selected voltage for Ulim.
To increase voltage:Turn clockwise
To decrease voltage: Turn counterclockwise
The selected voltage is now the maximum voltage value to which
Uset can be adjusted manually.
Setting Range
The <Ulim> potentiometer can be turned 270°, which
corresponds to a setting range of 0 V
Ulim Unom.
The 10 revolutions which are possible with the <Uset> rotary knob
[8] always correspond to a range of 0 V to Ulim.
4.1.3 Uset – Voltage Setpoint Value
Function
Predefined voltage at which the consuming device is to be
operated
Adjusting Uset
Uset is adjusted with the left-hand rotary knob [8] <Uset>.
First press the key [9] <DISPLAY Uset/Iset>.
The left-hand display [4] <Volt> is switched from the presently
measured voltage value Uout to the manually adjusted voltage
setpoint Uset.
Hold this key depressed.
At the same time, turn the rotary knob [8] <Uset>.
To increase voltage:Turn clockwise
To decrease voltage:Turn counterclockwise
If the output is activated <OUTPUT> “ON”, output voltage is
adjusted directly by turning rotary knob.
If the output is deactivated <OUTPUT> “OFF”, no voltage is present
at the output during adjustment. Voltage is not applied to the
output until it is activated <OUTPUT> “ON”.
After releasing the key [9] <DISPLAY Uset/Iset>, the display [4] is
switched back to the measured value for output voltage Uout.
CAUTION!
The Uset value can also be changed when the KONSTANTER is not in
operation.
Setting Range
The rotary knob can be turned 10 revolutions.
The 10 revolutions which are possible with the <Uset>
potentiometer [8] always correspond to a range of 0 V to Ulim.
The selection of a low value for Ulim allows for “finer” adjustment
of the Uset value.
The setting accuracy specified in chapter 1.4.3 makes reference
to the respectively displayed setpoint value.
19.99 is the highest value which can be displayed for Uout and
Uset with 20 V devices.
The A-D converter generates an overflow display for values
beyond this limit: