Datasheet

SLP KONSTANTER Series SLP 120 / 240 / 320
Laboratory Power Supplies
6 GMC-I Messtechnik GmbH
Characteristic Values, 320 W Series
Unless otherwise indicated, all entries represent maximum values and are valid within an operating temperature range of 0 to 50° C, the
nominal power range and a line voltage range of 230 V ± 10% after a warm-up period of 30 minutes.
1) Indicated values are increased by a factor of approximately 1.2 within a mains input voltage range of –10% to –15%.
Description (abbreviated designation) SLP 320-32
Type 32 N 32 R 18
Nominal Output Data Voltage Setting Range 0 ... 32 V
Current Setting Range 0 ... 18 A
Continuous Power at Tu 40° C max. 320 W
Short-Time Rating for t < 90 s / Tu 25° C max. 430 W
Current Derating at Tu > 40° C – 0.5 A / K
Output Operating Characteristics
Overall Setting Tolerance at 23 ± 5° C
with Reference to 3½ Place Setpoint Display
including System Deviation for Load / Line
Voltage
Current
0.2 % + 150 mV
0.5 % + 50 mA
Static System Deviation
1)
at 100% Load Variation
1)
Voltage
Current
30 mV
40 mA
Static System Deviation
1)
at 10% Line Voltage Variation
1)
Voltage
Current
10 mV
20 mA
Residual Ripple
1)
Voltage (10 Hz ... 10 MHz)
Current (10 Hz ... 1 MHz)
30 mV
eff
50 mA
eff
Common-Mode Interference (10 Hz ... 1 MHz) 0.5 mA
eff
Settling Time (voltage)
with Sudden Load Variations of 10 to 90% I
nom.
Tol eran ce
Δ I = 80%
64 mV
200 μs
Under and Overshooting with
Sudden Load Variations of 50 A / ms
Δ I = 80% 400 mV
Response Time (voltage)
with Setpoint Change 0 100%
with Setpoint Change 100% 0
Tol eran ce
Open-Circuit / Nominal Load
Open-Circuit / Nominal Load
64 mV
1 ms / 1 ms
1 ms / 1 ms
Response Time (current)
with Setpoint Change 0 100%
with Setpoint Change 100% 0
Tol eran ce
Short-Circuit / Nominal Load
Short-Circuit / Nominal Load
180 mA
< 5 ms / < 5 ms
< 5 ms / < 5 ms
Measurement Value Display (3½ place)
Measuring Resolution Voltage
Current
100 mV
10 mA
Measuring Accuracy at 23 ± 5° C
with Reference to the respective Measurement Value
Voltage
Current
0.2 % + 120 mV
0.5 % + 40 mA
Protective Functions
Output Overvoltage Protection Threshold 40 ± 1 V
Protection against Pole Reversal – Load Capacity Continuous 20 A
Reverse Voltage Resistance Continuous 64 V
General
Power Supply
1)
at Nominal Load
in Standby Mode
at max. Short-Time Power
230 V~ +10 / 15 %
47 ... 63 Hz
Power Consumption 650 VA; 460 W
50 VA; 15 W
770 VA
at Nominal Load
Typical
Efficiency at Nominal Load
in Standby Mode
at max. Short-Time Power
> 69 %
Switching Frequency 200 kHz
Article Number K234A