Operating Instructions MAVOSPEC Spectrometer 1/03.
Thank you for selecting this state of the art GOSSEN product. We hope you are enjoying the easy operation, the high quality, precision and wide functionality. Please check that the spectrometer and accessories as detailed below are included in the carrying case. If the contents are incomplete, please contact your local dealer.
Contents Page Safety Precautions, Warnings, Notes 4 1 1.1 1.2 Preparation Overview Charging Instructions and Battery Care 7 7 2 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.7.1 2.8 2.8.1 2.8.2 2.8.3 2.8.4 2.8.5 2.8.
Safety Precautions Please read these safety precautions carefully before using your measuring device. This will help you to avoid damaging the product and prevent personal injury. This icon identifies important warnings which should be read in any case before the initial start-up of your GOSSEN product. Warnings In the event of malfunction, switch off the measuring device immediately.
Handle rechargeable batteries with care. Rechargeable and normal batteries may leak or explode if handled incorrectly. Please adhere to the following safety precautions: Make sure that the measuring device is switched off before removing or inserting rechargeable batteries. If the measuring device is used with a power supply, then, first of all, the power supply must be disconnected (pull the mains plug out of the electrical outlet).
Icon for separate collection of recyclable materials / hazardous waste in European countries This icon indicates that this product must be disposed of separately. The following must be observed by users in European countries: This product may only be disposed of separately at a designated collection point. It may not be disposed of with household trash. For further information contact your local dealer or waste disposal authorities.
1 Preparation 1.1 Overview Power / Measure / Back button Changeable measuring head Press 1s to turn device on Press 8s to turn device off in emergency situations hard power off Turn to the left to remove the measuring head or to the right to attache one. During operation button has back or measurement functionality, depending on the selected menu. Touch screen display Displays the measurement results and allows the control of the device.
2 MAVOSPEC Software The MAVOSPEC software allows you to make measurements of light sources, store and present them in a tabular or graphic form and calculate chromaticity parameters. It works as an Android application and its use is similar to other applications for phones and devices based on the Android system. 2.1 Home Screen On the home screen, you can choose one of the following functions, which will be described further on.
2.2 Config Screen The following measurement parameters can be set on the config screen: Auto Device automatically selects integration time. When this field is checked, the integration time parameter is omitted. Integration time Duration of the light measurement. Repeat count Number of measurements which will be used to calculate average light signal. Flicker frequency Flicker frequency of a measured light source in Hz. Set 0 for non-flicker light source.
2.3 Spectrum Screen The Spectrum screen presents the spectrum chart of a measurement with several parameters at the top. The name of the measurement being presented is always shown in the title bar next to the GOSSEN icon. Screens which present measurements are equipped with a quick navigation tool bar. The quick navigation toolbar allows you to navigate between measurements in the current directory and select another presentation screen.
2.4 Results Screen The Results screen presents calculated chromaticity parameters which describe a measured light source. 2.5 CIE Chart Screen CIE chart screen shows chromaticity parameters on CIE chromaticity diagrams.
2.6 Data Screen The Data screen presents a table of spectrum data corresponding to a measurement. First column shows the wavelength, the second column the signal level. 2.7 Library Screen The Library screen shows measurements in the current directory and allows selecting the measurement which is displayed on the presentation screens. After a selection is made (by tapping the desired line), the measurement is displayed on the spectrum screen and is set as the current one.
When you tap a measurement on the list and hold it for a while, the context menu is displayed. The Context menu provides you with a set of functions available for a selected measurement. 2.7.1 Directories Screen The directories screen shows directories in which the measurement results are stored. It also provides you with any easy tool to manage them. They are created automatically on the micro SD card where the measurements are collected and named using the date on which they were taken.
2.8 Settings Screen System settings as data, time or Wi-Fi configuration can be edited in the Settings screen. 2.8.1 Battery Status Battery status shows the current battery level. Battery level is also shown below title bar as a line. Depending on the battery level it has color green (high level) , yellow (medium level) or red (low level). 2.8.2 Wi-Fi Settings For activating WiFi tap on the related box.
Locate your wireless network name (SSID). Tap on your wireless network's SSID. If your wireless router requires a key or password, the prompt on the right side should appear and you should enter the password. i Some keys/passwords are case-sensitive, so if it fails to connect, please make sure you use uppercase and lowercase letters appropriately. If you are successful, it should show the word Connected underneath your wireless router's SSID on this list.
2.8.2 Idle Time Idle time can be set to turn off the device automatically after the selected time and to lengthen operating time for battery. If idle time is off the device works until manual turn off. 2.8.3 Date & Time Settings Several settings for time and date as well as the formats can be done in this screen.
2.8.4 Status This screen shows the status of the last measurement. 2.8.5 Info Device information is shown in this pop up menu.
3 Trigger Input / Output MAVOSPEC is equipped with input and output signal lines dedicated for starting measurement from external device and triggering external devices when measurement is started. Trigger In line - high state on this line causes the same effect as pressing button on the device. Trigger Out line - at the measurement start MAVOSPEC sets high state on this line. The high state is maintained throughout the period of measurement.
Pin allocation – stereo mini jack plug 3.5 mm 1. Sleeve - GND 2. TRIGGER OUT - Open collector (Optocoupler TLP 127 - Darlington transistor). - Maximum Collector current: 50 mA. - The transistor is turned on during the measurement 3. TRIGGER IN - The measurement is triggered by the rising edge of voltage (3 - 9V) 4.
Possible wiring diagrams: MAVOSPEC MAVOSPEC GOSSEN Foto- und Lichtmesstechnik GmbH 20
4 Accessories The MAVOSPEC can be combined with different accessories for various measurement purposes. The device is equipped with a coder for the automatic detection of an attached accessory and for automatic calibration file download. When you purchase your MAVOSPEC with accessories, factory spectral calibration is included in the price. Simply remove the standard measuring head and you can attach any add-on to the spectrometer.
4.2 Display Sensor – MAVOPROBE 1.0 (M602A) MAVOPROBE 1.0 is an MAVOSPEC accessory luminance measurement of flat LCD, OLED panels and plasma FPDs. It is also a perfect solution for testing projection displays as well as for OLED light source measurement. The set includes an optical measuring probe, fiber optics and adaptor for the mini– spectrometer with a coder for automatic detection of accessory calibration file. The measuring probe can be installed on a tripod for distance measurements.
4.3 Software – GL SPECTROSOFT BASIC (M603A) GL SPECTROSOFT BASIC is an operation and evaluation software for MAVOSPEC with standard functionalities. The installation program for the software GL SPECTROSOFT is part of the CD in the standard delivery. The license has to be purchased separately and is copy protected by an USB dongle. GL SPECTRASOFT BASIC standard functionality: Spectral power distribution – graphic display of the spectrum with scaling function and assignment of colors to wavelengths.
Software – GL SPECTROSOFT PRO (M603B) 4.4 GL SPECTROSOFT PRO is an operation and evaluation software for MAVOSPEC with an enhanced range of functions. It includes all of the GL SPECTROSOFT BASIC functions. The installation program for the software GL SPECTROSOFT is part of the CD in the standard delivery. The license has to be purchased separately and is copy protected by an USB dongle.
5 Calibration The MAVOSPEC with intuitive user interface is one of the most accurate and reliable spectrometers in its class, and reflects the most up-to-date technology available on the market. Before shipment, each spectrometer is photometrically and radiometrically calibrated together with any optional ordered accessories.
6 Service and Maintenance No special maintenance is required, if the MAVOSPEC is handled correctly. Keep the outside surfaces clean. Use a slightly dampened cloth for cleaning. Do not use cleansers, abrasives or solvents. Should the device not work to your satisfaction, please send it to: GOSSEN Foto- und Lichtmesstechnik GmbH Lina-Ammon-Str.22 I 90471 Nürnberg I Germany Phone: +49 911 8602-181 I Fax: +49 911 8602-142 I E-Mail: info@gossen-photo.de www.gossen-photo.
7 Technical Specifications Photometry Application Illuminance [lux] Illuminance class Daylight, LEDs, halogen and more 1 lx … 200 000 lx Class B – DIN 5032-7 Class AA – JIS C 1609-1:2006 Error limit – cos like rating (f2') Distance, diffuser to surface to be measured Luminance [cd/m²] 1.9% 29 mm 1 cd/m² ... 10 000 cd/m², optionally with MAVOPROBE 1.
Interfaces and Memory USB USB 2.0 WiFi 802.11b/g Bluetooth SD card slot In preparation Micro SD Measured value memory Data format Automatic / 4 GB micro SD XML Plug for fiber-optic cable Optional SMA905D Operation Display Operation 3.
Printed in Germany – Subject to change without notice GOSSEN Foto- und Lichtmesstechnik GmbH I Lina-Ammon-Str.22 I D-90471 Nürnberg I Germany Telefon: +49 911 8602-181 I Fax: +49 911 8602-142 I E-Mail: info@gossen-photo.de www.gossen-photo.