™ AltaDyne AutoFloat Alternating Pressure Management Plus Model # 753003 Model # 753004 OPERATORS MANUAL Graham-Field Health Products 2935 Northeast Parkway Atlanta, Georgia 30360 1-800-347-5678 Fax: 1-800-726-0601 www.grahamfield.
Lumex® AltaDyne AutoFloat Warranty Table Of Contents 1.0 Warranty························································································································· 2 2.0 Symbols ·························································································································· 3 3.0 Indications for Use········································································································ 3 4.
Lumex® AltaDyne AutoFloat Indications for Use/Safety Precautions 2.0 Symbols Attention, consult accompanying document Dangerous voltage Type BF equipment Protective earth 3.0 Indications for Use This device is intended to assist in treating and preventing pressure ulcers. 4.0 Contraindication Air support treatment is not recommended when spinal stability is a concern. 5.0 Safety Precautions Review the following SAFETY PRECAUTIONS prior to using the AltaDyne Auto Float™ System.
Lumex® AltaDyne AutoFloat Description/Features 6.0 Description The AltaDyne AutoFloat is a portable air flotation system with on-demand alternating low pressure treatment, designed to provide benefit to patients suffering from, or at risk of, developing pressure ulcers. The system, consisting of a Control Unit and Air Flotation Mattress with it’s Top Sheet, is designed to provide pressure management and patient comfort.
Lumex® AltaDyne AutoFloat Operator Control Panel 8.0 Operator Control Panel 4 6 Auto Float APM Plus 5 3.2 3.4 7 3.3 9 8 3 3.1 1. ON/STANDBY switches Control Unit between On and STANDBY modes. 10 2 1 4. SOFT (–) decreases the inflation level of the bed while Control Unit is in manual mode only. Pressing once will decrease the setting by one comfort level. Pressing and holding will rapidly change the setting. 2. POWER INDICATOR LED indicates AC power is present to the Control Unit. 3.
Lumex® AltaDyne AutoFloat Setting Up the System/Instructions for Use 9.0 Setting Up the System Mattress Mattress 1. Place the mattress on the bed frame with the connecting hoses at the foot of the bed. 2. Before securing the mattress straps, raise the head and foot ends of the bed frame. Raise the side rails. Loosely, secure mattress to bed frame by looping the nylon straps on mattress through the bed frame and then back through the D-rings on mattress.
Lumex® AltaDyne AutoFloat Instructions for Use/Transferring Patients If the patient sinks into the mattress for any reason, (such as elevation of the head section or other part of the bed) the Control Unit will automatically increase the inflation level until the appropriate flotation height is again realized. Disconnecting the mattress sensor cord from the Control Unit while in automatic mode will cause an audible alarm to sound and indicators adjacent to the ALARM SILENCE and AUTO icons to flash.
Lumex® AltaDyne AutoFloat CPR/Cleaning/Care/Storage 12.0 CPR (Cardio Pulmonary Resuscitation) To deflate mattress for CPR: 1. Disconnect the dual quick-connect couplings of the mattress hose from the Control Unit. Mattress will deflate in approximately 15-20 seconds. 2. Proceed with CPR procedures. 13.0 System Cleaning, Care and Storage 13.1 Cleaning WARNING Disconnect the AC power cord from the wall outlet before attempting to clean the Control Unit.
Lumex® AltaDyne AutoFloat Troubleshooting/Specifications 14.0 Troubleshooting SYMPTOM: CORRECTIVE ACTION Mattress does not inflate or is too soft. Make certain the air hoses are not kinked, cracked, or split. Verify the hose couplings are fully engaged to controller. Verify that all 20 air cells are connected to the air manifold within mattress. Power loss indicator flashing and audible alarm sounding Power has been disconnected from unit while in an operational mode.