Owner's manual
OLS200, CSG200     15214 Ver 9 - January 2010 57U 
Operating Manual  Page 12  www.grant.co.uk 
7.3  Replacement of fuses 
  WARNING: Before removing the cover, isolate from the mains power supply. 
  Only a competent person should change fuses. 
7.3.1  OLS200 Shaker bath 
  First disconnect the mains cable from the power supply, then remove the socket end from the 
plug in the back of the bath. Drain the bath. Press down the fuse drawer catch (see figure 12). 
  Pull out the fuse drawer, replace the fuse with the correct type, and replace the fuse holder. 
  The fuses are Littelfuse 3AB 314 series, fast-acting, high breaking current (max breaking current 
at least 750 A); dimensions are 1.25 inch long, 0.25 inch diameter. Replace fuses only by the 
same type and rating. 
  230V units – 10A (250V) 
  120V units – 15A (250V) 
7.3.2  Refrigerated cooler: CS200G 
  First disconnect the mains cable from the power supply, then remove the socket end from the 
plug in the back of the unit. Press down the fuse drawer catch (see figure 12). 
  Pull out the fuse drawer, replace the fuse with the correct type, and replace the fuse holder. 
  The fuses are 1.25 x 0.25 inch ceramic quick acting, rated:  
  230V units – 10A (250V) 
  120V units – 20A (250V) 
8  Guarantee 
  When used in laboratory conditions and according to these instructions, this equipment is 
guaranteed for three years against faulty materials or workmanship. 
9  Service 
  For service, return for repair to our Service Department in the UK or, in other countries, to our 
distributor. 
    Service Address: 
Grant Instruments (Cambridge) Ltd., 
Service Department, 
Shepreth, 
Cambridgeshire 
SG8 6GB 
England 
Tel:    +44 (0) 1763 260811 
Fax:   +44 (0) 1763 262410   