Operating Manual
GLS Aqua Plus    29339 Ver. 3  
Operating Manual    Page 12 
  www.grant.co.uk 
EN 
7.  Press the enter key to confirm the value and the bath will display  ait for the 
bath to return to normal mode. 
High calibration point ( ) 
1.  Set the bath to a temperature between 70°C and 90°C, allow the bath  to stabilise, 
indicated by the display not fluctuating by more than 0.2°C. 
2.  Measure  the  temperature  of  the  water  with  a  calibrated  thermometer  and  make  a 
note of this reading. 
3.    
4.  -navigate to   
5.  - navigate to   
6.  -
reading taken earlier. 
7.  Press the enter key to confirm the value and the bath will display  ait for the 
bath to return to normal mode. 
Setting the calibration ( ) 
1.  Press the enter key. The display will   
2.  -navigate to   
3.  - navigate to  apply the new 
low and high calibration values. 
The bath is now calibrated and the display will read the corrected value. 
Restoring factory calibration ( ) 
1.  Press the enter key. The display will   
2.  -navigate to  press the enter key. 
3.      -            
factory settings. 