3.5.  Bath setup 
The key settings on the Optima can be changed by selecting Bath from the Setup option on 
the main control window menu. 
3.5.1.  On timer completion 
This section allows the timer period to be set and defines the behaviour when the countdown 
timer reaches zero. This behaviour is only applicable to the GR150 and GP200 models. 
3.5.2.  Presets 
This section allows the set temperature and pump speed to be configured for the three 
presets. Pump speed configuration is only available with the TXF200 model. 
3.5.3.  Alarms 
This section allows alarms to be set during normal operation. A high and/or low alarm can 
be set up. There are three types of alarm: 
  Disabled – No alarm 
  Fixed – Alarm triggered at a fixed temperature 
  Deviation – Alarm triggered when the deviation from the set temperature is exceeded 
The alarm temperature defines either a fixed temperature or a deviation temperature 
depending on the alarm type. The checkboxes define the actions taken if an alarm is 
triggered. The Buzzer and Relay 1 can be switched on and latched if a persistent alarm is 
required. Relay 2 is only available with a GP200. Alarm Hold Off specifies the time before 
the alarm actions are retriggered once an alarm has been acknowledged. Alarms can be 
acknowledged by clicking on the temperature alarms box. The TX150/TXF200 user manual 
has further information on the implementation and behaviour of the alarms.   