Manual
XUBA Series    29667 V1 
Operating Manual 
4 
Fill Line 
Quick guide 
Pour cold water into the bath so that it reaches the 
minimum tank fill line (but does not go over the 
maximum fill line). 
Add the required dose of detergent (see page 10). 
Operate the  ultrasonics by  turning  the timer  dial  to 
the required  time and pressing the SONICS button 
(where applicable). 
Operate the heater, if the  bath has a heating func-
tion, by turning the heater dial to the required tem-
perature and pressing the HEAT button. 
Turn on the  ultrasonics for approximately  5-10  min-
utes  to  degas  the  cleaning  liquid.  (See  page  8  for 
more details on degassing). 
The bath is now ready to use. 
At the end of the cycle, remove the basket from the 
bath and rinse the items under clean running water. 
The lid can be inverted and the basket placed on top 
to catch excess fluid as the items dry. 
Please refer to the following pages in this manual for more detailed instructions. 
Change the cleaning liquid at regular intervals. Your cleaning process will determine 
how often to change the liquid  the more soiled your items, the more often you will 
need to change the liquid.  
Grant recommends changing the cleaning liquid at least on a daily basis. 