12 Operation Of Logger
The device supports two different logger functions:
„Func-Stor“: manual recording via keypress “Store” (key 6).
„Func-CYCL“: automatic recording with the selected recording interval/cycle
The logger records 1 measurement result each time
The data set consists of: - measuring value at time of recording
- time and date of recording
For the read out and evaluation of the data the software GSOFT3050 (V1.7 or higher) has to be
used. The software also allows easy configuration and starting of the logger.
When the logger is activated (Func Stor or Func CYCL) the hold function is no longer available, key
6 is solely used for the operation of the logger functions.
12.1 Storing Single Measurements („Func-Stor“)
a) Record Measuring:
If the logger function “Func Stor” was activated (see chapter 8 “Device Configuration”), up to 99
measuring data sets can be recorded manually.
Press shortly: Data set will be recorded (“St. XX” will be shown shortly. XX is the
number of the data set)
If the logger memory is full, the display will show:
b) Viewing Recorded Measurings:
Stored data sets can be either read out via PC Software GSOFT3050, or be viewed in the
display directly.
Press for 2 seconds: The display will show:
Please note: ”rEAd Logg” only appears, if there are already data stored! If memory is
empty, the configuration menu will show.
Press shortly: Changing between measuring value and date+time of data set
or
Changing between the data sets
Exit logger data display
c) Clear recorded measurings:
If there are already data sets stored, these can be deleted via the store key:
Press for 2 seconds: Calling of clear-Menu
Change selection by: or :
Clear nothing (cancel menu)
Clear all recordings
Clear the last recording
Stores the settings