2 Device Configuration
For configuration of the device press "Set"-key (key 4) for 2 seconds, the first menu will be shown.
Choose between the individual values that can be set by pressing the "Set"-key (key 4) again.
The individual values are changed by pressing the keys "5" (key 2) or "6" (key 5).
Use key "Store" (key 6) to leave configuration and to store settings.
2.1 'Type': Selection of the Thermocouple-Type
ni.cr: type K, NiCr-Ni
n: type N, NiCrSi-NiSi
S: type S, Pt10Rh-Pt
t: type T, Cu-CuNi
J: type J, Fe-CuNi
2.2 'Resolution': Selection of Display Resolution (Not for Type S!)
1°: Resolution 1°C
0.1°: Resolution 0.1°C
Auto: Resolution is selected automatically
Attention: When type S is selected the resolution is always 1°C!
2.3 'Unit': Selection of Temperature Unit °C /°F
°C: All temperature values in degrees Celsius
°F: All temperature values in degrees Fahrenheit
2.4 'Corr': Selection of Display Correction Factor
0.950...1.200: The temperature value (referring to 0°C or 32°F) will be multiplied by this factor.
Thus, in case of surface measurements, losses by transfer can be compensated for (>1.000).
Furthermore the factor can be used to compensate probe deviations.
off: Factor deactivated (=1.000)
2.5 'Offset': Zero Displacement
-10.0°C...10.0°C The zero point of the measurement will be
or displaced by this value so as to compensate
-18.0°F...18.0°F: for deviations in the sensor and in the measuring unit.
off: Zero displacement deactivated (=0.0°)
2.6 'Power.off': Selection of Power-Off Delay
1...120: Power-off delay in minutes.
Device will be automatically switched off as soon as this time has elapsed if no key is
pressed/no interface communication takes place.
off: Power-off function deactivated (continuous operation, e.g. in case of mains operation)
2.7 'Out': Function of the Output
off: No output function, lowest power consumption
SEr: Output is serial interface
dAC: Output is analogue output 0...1V
2.8 'Address': Selection of Base Address when Output = Serial Interface
01, 11, 21, ..., 91: Base address for interface communication.