User manual
H61.0.01.6C-03 User’s Manual GMH3156 page 4 of 12
_______________________________________________________________________________________________________________________
GREISINGER electronic GmbH Hans-Sachs-Straße 26 Fon: 0049 - 9402 / 9383-0 http://www.greisinger.de
D-93128 Regenstauf Fax: 0049 - 9402 / 9383-33 eMail: info@greisinger.de
2 Configuration
To change device settings, press Menu (key 4) for 2 seconds. This will activate the configuration menu (main
display: „SEt“).
Pressing key Menu changes between the menus, pressing 4(key 3) jumps to the referring parameters,
which can be selected with key 4(key 3).
The parameters can be changed with 5 (key 2) or 6 (key 5).
Pressing Menu again jumps back to the main configuration menu and saves the settings.
Quit (key 6) finishes the configuration and returns to standard measuring operation.
Menu Parameter Values Meaning
KEY
Menu
KEY
4
44
4
KEY
5
or
6
Set Configuration: Generic Configurations
Unit
mbar,bar..
Unit: Unit of display (given by sensor 1 when using 2 sensors) *,**
SL
oFF/on
Sea level correction: on or off (only for Sensor 1) *,**
Alti
-2000..9999
Altitude above sea level [m] (only for Sensor1 and if SL ) *,**
Rate: Measuring rate (p.r.t. chapter 2.1) *
Slo
Slow measuring rate (4Hz filtered, low power consumption) *
FASt
Fast measuring rate, filtered (>1000Hz) *
rAtE
P.dEt
Peak detection: fast measuring rate, unfiltered (>1000Hz) *
1-120
Averaging period in seconds, used by the averaging function
t.AVG
oFF
Averaging function deactivated
1-120
Auto Power Off time in minutes
P.oFF
oFF
Auto Power Off deactivated
oFF
Function of the output: No output function, lowest power consumption
SEr
Output is serial interface
Out
dAC
Output is analogue output 0...1V
Adr.
01,11..91
Base address of interface (if Out = SEr)
dAC.
CH1, CH2
or CH DIF
Choice of the input to be the source for the analogue output
(if Out = dAC)
dAC.0
eg.
-5.00..5.00 mbar
Enter desired value at which the analogue output potential should be 0V
(if Out = dAC)
SEt
ConF
dAC.1
eg.
-5.00..5.00 mbar
Enter desired value at which the analogue output potential should be 1V
(if Out = dAC)
SEt
Set Calibration: Ad
j
ustment of Sensors
Sensordep., e.g.
-5.00..5.00 mbar
The offset of sensor 1 will be displaced by this value to compensate for
deviations in the probe or in the measuring device.
OFS.1
OFF
Zero displacement inactive (=0.0°)
-2.000...2.000
The measuring scale of sensor 1 will be changed by this factor [%] to
compensate deviations of temperature probe or measuring device
SCL.1
OFF
Scale correction factor inactive (=0.000)
Sensordep., e.g.
-5.00..5.00 mbar
The offset of sensor 2 will be displaced by this value to compensate for
deviations in the probe or in the measuring device.
OFS.2
oFF
Zero displacement inactive (=0.0°)
-2.000...2.000
The measuring scale of sensor 2 will be changed by this factor [%] to
compensate deviations of temperature probe or measuring device
CAL
SCL.2
oFF
Scale correction factor inactive (=0.000)
SEt
Set Alarm: Settin
g
s Of Alarm Function
AL. AL. 1
On
Alarm sensor 1 on, with buzzer sound
no.So
Alarm sensor 1 on, without buzzer sound
oFF
no alarm function for sensor 1
AL.Lo/AL.1
Sensor1-Min...
AL.1-Hi
Min alarm rail Sensor 1 (not when AL.1 = oFF)
Sensor1-Min is the lower display range of sensor 1
AL.Hi/AL.1
AL.1-Lo...
Sensor1-Max
Max alarm rail Sensor 1 (not when AL.1 = oFF)
Sensor1-Max is the upper display range of sensor 1
AL. 2
On
Alarm sensor 2 on, with buzzer sound
no.So
Alarm sensor 2 on, without buzzer sound
oFF
no alarm function for sensor 2
AL.Lo/AL.2
Sensor2-Min...
AL.2-Hi
Min alarm rail Sensor 2 (not when AL.2 = oFF)
Sensor2-Min is the lower display range of sensor 2