User manual

H61.0.01.6C-03 User’s Manual GMH3156 page 9 of 12
_______________________________________________________________________________________________________________________
GREISINGER electronic GmbH Hans-Sachs-Straße 26 Fon: 0049 - 9402 / 9383-0 http://www.greisinger.de
D-93128 Regenstauf Fax: 0049 - 9402 / 9383-33 eMail: info@greisinger.de
! fast measuring rates (rAtE FASt,P.dEt):
- sensor 1, arithmetic mean value since the last recording
- sensor 1 min peak, max peak since the last recording
- sensor 2, arithmetic mean value since the last recording
- sensor 2 min peak, max peak since the last recording
- difference sensor 1 – sensor 2, arithmetic mean value since the last recording
- difference sensor 1 – sensor 2, min peak, max peak since the last recording
Starting a recording:
By pressing "Store" (key 6) for 2 seconds the logger operation will be called. The display will show:
By pressing "Store" again the recording will be initiated.
After that the display shows ‘St.XXXX‘ for a short time whenever a measuring is recorded. XXXX is the
number of the measuring 1..4000.
If the logger memory is full, the display will show:
The recording automatically will be stopped.
If Low-Power-Logger-Function „Lo.Po = on“ the device switches itself off as soon as the memory gets filled.
Stopping the recording manually:
By pressing "Store" (key 6) the recording can be stopped manually. Then the following choice appears:
The selection can be made by 5 (key 2) and 6 (key 5). "Quit" (key 6) enters the choice.
Note: If you try to switch off the instrument in the cyclic recording operation You will be asked
once again if the recording is to be stopped.
The device can only be switched off after the recording has been stopped!
The Auto-Power-Off-function is deactivated during recording!
Clear Recordings:
By pressing "Store" (key 6) for 2 seconds the logger operation will be called.
The display will show:
By pressing the keys 5 (key 2) or 6 (key 5) the display will change to .
When „Store“ is pressed, the choice for clearing the logger memory will be displayed:
Clear all recordings
Clear the last
recording sequence
Clear nothing
(cancel menu)
The selection can be made by 5 (key 2) and 6 (key 5). "Quit" (key 6) enters the choice.
Stop the
recording
Do not stop the
recording