8 Display Functions
8.1 Measuring display
During normal operation the temperature in [°C] or [°F] is displayed.
8.2 Min/Max Value Memory
watch Min values (Lo): press shortly once display changes between „Lo„ and Min values
watch Max values (Hi): press shortly once display changes between „Hi„ and Max values
restore current values: press or once again current values are displayed
clear Min-values: press for 2 seconds Min values are cleared. Display shows shortly „CLr„.
clear Max-values: press for 2 seconds Max values are cleared. Display shows shortly „CLr„.
After 10 seconds the currently measured values will be displayed again.
9 Error and system messages
Display
Description
Possible fault cause
Err.1
Measuring range
exceeded
Wrong signal
Err.2
Measuring value below
measuring range
Wrong signal
Err.7
System fault
Error in device
Err.9
Sensor error
Sensor or cable defective
Er.11
Calculation not possible
Calculation variable
missing or invalid
8.8.8.8
Segment test
The transducer performs a display test for 2 seconds after power
up. After that it will change to the display of the measuring.