10 Configuration of the device
Follow these instructions to configure the functions of the device:
Press SET until the first parameter UNIT appears in the display
If a parameter should be changed, press or ,
The device changed to the setting edit with or
Confirm the value with SET
Jump to the next parameter with SET.
Parameter
value
information
SET
and
UNIT
Unit of temperature displays factory setting: °C
°C
Temperatures in °Celsius
°F
Temperatures in °Fahrenheit
VT
Display at zero output of humidity display value (output scaling)
-xxx.0 … +xxx.0
Entry of display value at which the temperature output should deliver
4 mA (or 0V) (factory setting: 0.0)
VT
Display at maximum output for humidity display value (output scaling)
-xxx.0 … +xxx.0
Entry of display value at which the temperature output should deliver
20 mA (or 1 V / 10 V) (factory setting: 100.0)
Offset correction of humidity measuring *)
oFF
deactivated (factory setting)
-5.0 … +5.0
Selectable from -5.0 to +5.0 % rel. humidity
Scale correction of humidity measuring *)
oFF
deactivated (factory setting)
-15.00 … +15.00
Selectable from -15.00 to +15.00 % scale correction
Pressing SET again stores the settings, the instruments restarts (segment test)
Please note: If there is no key pressed within the menu mode within 2 minutes, the
configuration will be cancelled, the entered settings are lost!
*) if higher values are needed, please check sensor, if necessary return to manufacturer for inspection.
Calculation: corrected value = (measured value Offset) * (1+Scale/100)
ATTENTION!
If the jumper is removed from
the shown contacts, the
configuration is inaccessible,
values are protected.
Never connect other contacts!