----: no measurement channel allocated
Indicates, that the displayed channel is configured with no allocated EASYBUS-measurement chan-
nel.
Possible causes: - incorrect configuration
Remedies: - perform new configuration of the EB3000
t.out: Timeout occurrence
This error message alternates with the current display value and indicates, that’s not possible to
enquire the EASY
BUS-measurement channel within the minimum required updating-rate.
Possible causes: - a system initialisation was performed
- incorrect configuration
- bus is blocked in fact of a master requiry
Remedies: - remove alarm by pushing key 4 or by using EASYBUS-Configurator
software
- perform new configuration of the EB3000
The following system-error-codes are defined:
. : no measurement channel for indication selected
(only a point is indicated down right of the display)
Indicates, that all channels are configured with display endurance 0.
Er.49: Capacity overload of the EASYBUS-output
This error message alternates with the current display value and indicates, that the max. allowable
value of the EASY
BUS-output has been exceeded.
Possible causes: - too many EASY
BUS-sensor modules connected on EB3000
- connection line with shortcut
Remedies: - check connection of the EASY
BUS-sensor modules