User manual

T20.0.2X.6C-05 page 4 of 4
To configure the parameters proceed like follows:
1. Press the key 1 (SET) for more than 4 sec’s until 'Unit' with temperature arrow appears in the display.
I.) 'Unit' with Temp-arrow: Temperature unit
Enter the desired Temperature unit. All referring settings and displays are done in this unit.
2. Choose the desired unit by pressing 2 (down) or 3 (up) key. Choice between °C and °F (ex works: °C)
3. Enter by pressing key 1 (SET), select next parameter by pressing key 1 again: 'OFFS' without temperature arrow
appears in the display.
II.) 'OFFS': Offset of humidity measuring (correction of measuring deviations):
The offset of the measuring will be shifted by this value, the input is in %RH. Calculation: see above.
4. Choose the desired value by pressing 2 (down) or 3 (up) key.
Max. input range: -5.0...5.0 %RH or 'oFF': offset is deactivated (oFF=0.0%, ex works)
5. Enter by pressing key 1 (SET), select next parameter: 'SCAL' without temperature arrow appears in the display.
III.) 'SCAL': Scale of humidity measuring (correction of measuring deviations):
The scale of the measuring is changed by this value. Calculation: see above.
6. Choose the desired value by pressing 2 (down) or 3 (up) key.
Max. input range: -5.00...5.00 or 'oFF': scale is deactivated (oFF=0.00, ex works)
7. Enter by pressing key 1 (SET), select next parameter: 'OFFS' with temperature arrow appears in the display.
IV.) 'OFFS': with Temp-arrow: Offset of temperature measuring (correction of measuring deviations):
The offset of the measuring will be shifted by this value, the input is in °C. Calculation: see above.
8. Choose the desired value by pressing 2 (down) or 3 (up) key.
Max. input range: -5.0...5.0 °C / -9.0...9.0°F or 'oFF': offset is deactivated (oFF=0.0, ex works)
9. Enter by pressing key 1 (SET), select next parameter: 'SCAL' with temperature arrow appears in the display.
V.) 'SCAL': with Temp-arrow: Scale of temperature measuring (correction of measuring deviations):
The scale of the measuring is changed by this value. Calculation: see above.
10. Choose the desired value by pressing 2 (down) or 3 (up) key.
Max. input range: -2.00...2.00 or 'oFF': scale is deactivated (oFF=0.00, ex works)
11. Enter by pressing key 1 (SET), select next parameter: 'AL.Lo' without temperature arrow appears in the display.
VI.) 'AL.Lo': min. alarm-point for humidity-measuring:
Enter the desired min. alarm-point. At resp. below the value that has been set an alarm message will occur.
12. Choose the desired value by pressing 2 (down) or 3 (up) key. Max. input range: 0.0 to AL.Hi
13. Enter by pressing key 1 (SET), select next parameter by pressing key 1 again: 'AL.Hi' without temperature arrow
appears in the display.
VII.) 'AL.Hi': max. alarm-point for humidity-measuring:
Enter the desired min. alarm-point. At resp. above the value that has been set an alarm message will occur.
14. Choose the desired value by pressing 2 (down) or 3 (up) key. Max. input range: AL.Lo to 100.0
15. Enter by pressing key 1 (SET), select next parameter: 'AL.dE' without temperature arrow appears in the display.
VIII.) 'AL.dE': alarm-delay for humidity-measuring:
The alarm will start, when the alarm condition exists for the entered delay time.
16. Choose the desired value by pressing 2 (down) or 3 (up) key. Max. input range: oFF; 0...9999 min.
17. Enter by pressing key 1 (SET), select next parameter: 'AL.Lo' with temperature arrow appears in the display.
IX.) 'AL.Lo' with temperature arrow: min. alarm-point for temperature-measuring:
Enter the desired min. alarm-point. At resp. below the value that has been set an alarm message will occur.
18. Choose the desired value by pressing 2 (down) or 3 (up) key. Input range: -40.0 to AL.Hi °C (AL.Hi with temp.arrow)
19. Enter by pressing key 1 (SET), select next parameter: 'AL.Hi' with temperature arrow appears in the display.
X.) 'AL.Hi' with temperature arrow: max. alarm-point for temperature-measuring:
Enter the desired min. alarm-point. At resp. above the value that has been set an alarm message will occur.
20. Choose the desired value by pressing 2 (down) or 3 (up) key. Input range: AL.Lo...120.0 °C (AL.Lo with temp.arrow)
21. Enter by pressing key 1 (SET), select next parameter: 'AL.dE' with temperature arrow appears in the display.
XI.) 'AL.dE' with temperature arrow: alarm-delay for temperature-measuring:
The alarm will start, when the alarm condition exists for the entered delay time.
22. Choose the desired value by pressing 2 (down) or 3 (up) key. Max. input range: oFF; 0...9999 min.
Enter by pressing key 1 (set). After pressing 1 again, the instrument will restart, (display shows 8.8.8.8 segment test).
Notice: In case an alarm is active, the display will show alternating AL.Lo resp. AL.Hi (without resp. with temperature
arrow) and the measured value