Specification:
Display range: relative air humidity:  0.0...100.0 % rel. humidity (temperature compensated)
temperature:  -25.0...70.0 °C or -13.0...158.0 °F
Rec. range of application: standard: 20,0...80,0 %RH
option high humidity:  5,0...95,0 %RH
Accuracy: (at nominal temperature = 25°C)
relative air humidity:  ±1% linearity, 2% hysteresis 
(at recommended range of application)
temperature: 0.4% of measured value ±0.3°C
Sensors: relative air humidity: cap. polymer sensor
temperature: Pt1000
Temperature compensation: automatic
Min-/max-value memory: Min and max measured values are stored
Output signal: EASY
BUS-protocol
Connection: 2-wire EASY
BUS, polarity free
Busload: 1.5 EASY
BUS-devices
Adjusting:  via interface or keypress (only with option VO)
by input of offset and scale (humidity and
temperature)
Display: (optional) approx. 10 mm high, 4-digit LCD-display
Ambient conditions for electronics:
Nominal temperature: 25 °C
Operating temperature: -25 to 70 °C
Relative humidity: 0 to 95 %RH  (non-condensing)
Storage temperature: -25 to 70 °C
Housing: ABS
Dimensions: 70 x 70 x 26 mm
Mounting: With holes for wall mounting (in housing - accessible after cover has been removed).
Mounting distance: 60mm, max. shaft diameter of mounting screws is 4 mm.
Electrical connection: 2-pin screw-type terminal, max. wire cross section: 1.5 mm²
EMC: The
 device corresponds to the essential protection ratings established in the Regulations of the Council for the
Approximation of Legislation for the member countries regarding electromagnetic compatibility (2004/108/EG).
In accordance with 
EN61326 +A1 +A2 (appendix A, class B), additional errors: < 1% FS.
When connecting long leads adequate measures against voltage surges have to be taken.
GREISINGER electronic GmbH
D - 93128 Regenstauf, Hans-Sachs-Straße 26
Tel.: 0049 9402 / 9383-0, Fax: 0049 9402 / 9383-33, eMail: info@greisinger.de