5 Start Operation
Connect sensor, turn on device via key.
After segment test the device displays some configuration:
• If a zero point adjustment was carried out the display shows shortly „nuLL Corr“.
After that the device is ready for measuring.
6 Configuration
To change device settings, press Menu (key 4) for 2 seconds. This will call the configuration menu.
Pressing key Menu jumps between the parameters.
The parameters can be changed with (key 2) or (key 5).
Quit (key 6) finishes the configuration and returns to standard measuring operation.
Parameter
Values
Description
‚Menu‘
or
ni.cr ,
n , S , t , J
Type: Selection oft he Thermocouple-Type
ni.cr: type K, NiCr-Ni
n: type N, NiCrSi-NiSi
S: type S, Pt10Rh-Pt
t: type T, Cu-CuNi
J: type J, Fe-CuNi
ºC, ºF
Unit: Selection of Temperature Unit
1ºC, 0.1ºC,
Auto
Resolution: Selection of Display Resolution (Not for Type S)
Attention: When Type S is selected the resolution is always 1°C!
0.950...1.200
Display correction factor (refer to chapter 9.2)
off
Factor deactivated (=1.000)
1 … 120
Auto Power-Off time in minutes
oFF
Auto Power-Off deactivated
01, 11 … 91
Base address of interface
-10ºC...10ºC
bzw.
-18ºF...18ºF
The offset of sensor will be displaced by this value to compensate for
deviations in the probe or in the measuring device.
oFF
Zero displacement inactive (=0.00)
SEr, dAC
Output: function of device output: serial interface or analogue output
oFF
No output function, lowest power consumption
-220ºC...1372ºC
Output Offset when output = Analogue outeput
(depending on selected range) Input of the temperature at which 0V
should be output
-220ºC...1372ºC
Output grade when output = Analogue Output
(depending on selected range) Input of the temperature at which 1V
should be output