User manual
H49.0.02.6C-07 User’s Guide GMH 3750 page 7 of 16
_____________________________________________________ _____________________________________________________________________________
5 Start Operation
Connect Temperature probe, turn on device via key.
After segment test the device displays some configuration:
• If the logger function is not off the time of the integrated clock will shortly be displayed.
• If a zero point adjustment was carried out the display shows shortly „nuLL Corr“.
After changing the battery the clock-setting menu is activated automatically (‚CLOC‘). Check the clock
and adjust, if necessary (p.r.t. chapter 6).
After that the device is ready for measuring.
6 Configuration
To change device settings, press Menu (key 4) for 2 seconds. This will call the configuration menu (main
display: „SEt“ or “rEAd LoGG”).
Pressing key Menu changes between the menus, pressing ( (key 3) jumps to the referring parameters,
which can be selected with key ( (key 3).
The parameters can be changed with (key 2) or (key 5).
Pressing Menu again jumps back to the main configuration menu and saves the settings.
Quit (key 6) finishes the configuration and returns to standard measuring operation.
Menu(*1)
Parameter/Values(2*)
Meaning
p.r.t.
‚Menu‘
or
Read Logg: single value-logger read out
(only when data existing! Please refer to chapter 12)
8
Set Configuration: Generic Settings
°C:
All temperature values are in degrees Celsius
*
°F:
All temperature values are in degrees Fahrenheit
0.1° / 0.01°:
Resolution 0.1° / 0.01°
*
7.1
Auto:
Resolution is selected automatically
E.751
characteristic curve according to EN60751
*
7.4
USEr
User sensor curve (Predefined to EN60751 values,
changeable by software GMHKonfig)
-2.50...2.50°C /
-4.50...4.50°F
Zero correction
*
10.1
oFF:
Zero displacement inactive (=0.0°)
-2.000...2.000:
Scale correction [in %]
*
10.2
oFF:
Scale correction factor inactive (=0.000)
1...30:
Average filter (period in seconds)
*
7.2
oFF:
Average filter inactive
1...120
Power-off delay in minutes. Device will be
automatically switched off as soon as this time has
elapsed if no key is pressed or no interface
communication takes place
oFF
Power-off function inactive (continuous operation)
oFF:
No output function
9
SEr:
Output is serial interface
dAC:
Output is analogue output
01,11..91
Base address of device for interface communication
9.1
-200..850°C /
–328...1562°F
Zero point of analogue output: Temperature at
which the analogue output potential should be 0V
9.2
1 2 3
4 5 6
ON
OFF
max
Store
CAL
min
Set