User manual
H49.0.02.6C-07 User’s Guide GMH 3750 page 8 of 16
_____________________________________________________ _____________________________________________________________________________
-200..850°C /
-328...1562°F
Scale of analogue output : Temperature value at
which the analogue output potential should be 1V
9.2
Set Alarm: Setting of Alarm Function
On / No.So
Alarm on with horn-sound / Alarm on without horn-
sound
7.3
OFF
no alarm function
-200°C ...AL.Hi
Min alarm rail (not when AL. oFF)
AL.Lo ... 850°C
Max alarm rail (not when AL. oFF)
Set Logger: Setting of Logger Function
CYCL
Cyclic: logger function cyclic logger
*
8
Stor
Store: logger function individual value logger
OFF
No logger function
0:01... 60:00
Cycle time of cyclic logger [minutes:seconds]
*
8
Set Clock: Einstellen der Echtzeituhr
HH:MM
Clock: Setting of time hours:minutes
7.4
YYYY
Year
TT.MM
Date: day.month
(*) If the logger memory contains data already, the menus/parameters marked with (*) can not
be invoked! If these should be altered the logger has to be stopped and the memory has to
be cleared before!
Hint: Restoring of ex-works settings
The settings will be set to the settings ex works, if keys ‘Set‘ and ‘Store‘ are pressed simultaneously
for more than 2 seconds.
7 Remarks To Special Features
7.1 Display Resolution
Standard setting: 'Auto', i.e. the device automatically switches over to the optimum resolution between .01°
and 0.01°.
If temperatures to be measured are near the switching threshold, a fixed resolution may be better, e.g. for
easy manual recording. In such a case please set the optimum resolution to the desired value.
7.2 Average filter („t.AuG“)
The filter forms the arithmetic floating average over the entered time. The displayed value will be
correspondingly slower in this case.
7.3 Alarm („AL.“)
There are three possible settings:
Alarm off (AL. oFF), on with horn sound (AL. on), on without horn sound (AL. no.So).
Following conditions will display an alarm, when the function is activated (on or no.So):
- Value is below lower (AL. Lo) or above upper alarm rail (AL.Hi).
- Sensor error
- Low battery (bAt)
- Err.7: System error (always with sound)
In case of an alarm and when polling the interface the “prio”-flag is set in the returned message.