H49.0.02.6C-07 User’s Guide GMH 3750 page 10 of 16
_____________________________________________________ _____________________________________________________________________________
c) Clear recorded measurings:
If there are already data sets stored, these can be deleted via the store key:
Press for 2 seconds: Calling of clear-Menu
Change selection by: or :
Clear nothing (cancel menu)
Clear all recordings
Clear the last recording
Stores the settings
8.2 Automatic Recording With Selectable Cycle Time „Func CYCL“
If the logger function “Func CYCL” was activated (see chapter 6 Configuration), and the logger was started,
it automatically records data sets at intervals of the selected logger cycle time
The logger cycle time is selectable from 1s to 60min (see chapter 6 “Configuration”).
Max. number of measurings: 16384
a) Starting a recording:
press for 2 seconds: the recording will be started.
Each recodring is signales with a short diplay of ‘St.XXXXX’. XX is the number of the data set.
If the logger memory is full, the display will show
b) Stop the recording:
Press for 2 seconds: If a recording is running, the Stop menu will be shown
Change selection by: or :
Do not stop the recording
Stop the recording
Stores the settings
Note: If you try to switch off the instrument in the cyclic recording operation, You will be asked
automatically if the recording should be stopped.
The device can only be switched off after the recording has been stopped!
The Auto-Power-Off-function is deactivated during recording!
c) Clear recorded measurings:
Press for 2 seconds: If logger data are present and the recording was stopped, the
clear menu will be shown
Change selection by: or :
Clear nothing
(cancel menu)
Clear all recordings
Stores the settings