E44.X.00.6C-02 Connection and operating manual GIR 300 page 13 of 28
- Press button 1 again. The display will show “Li(Limit = Measuring range limit).
- Use button 2 and button 3 to select the desired measuring range limit.
Display
Measuring range limit
Notes
off
Deactivated
Exceeding of the measuring range limit is
tolerable for about 10% of the selected input
signal.
on.Er
Active, (displays error)
The measuring range limit is exactly bounded
by the input signal. When exceeding or short-
falling the input signal the device will display
an error message.
on.rG
Active, (displays the selected limit)
The measuring range limit is exactly bounded
by the input signal. When exceeding or short-
falling the input signal the device will display
the selected lower/upper display value.
[e.g. humidity: when shortfalling or exceeding,
the device will display 0% or 100%]
Hint: When exceeding the measuring range limit > 10% independently from the setting, the device
will always display an error message (“Err.1“ or “Err.2“).
- Press button 1 to validate the selection, the display shows “Li“ again.
- When pressing button 1 again, the display will show “FiLt“ (Filter = digital filter).
- Use button 2 and button 3 to select the desired filter [in sec.].
Selectable values: 0.01 ... 2.00 sec.
Explanation: this digital filter is a digital replica of a low pass filter.
Note: when using the input signal 0-50 mV a filter value of at least 0.2 is recommended
- Press button 1 to validate your value, the display shows “FiLt“ again.
Now your device is adjusted to your signal source. Now the only thing left to do is to adjust the outputs of the device.
- When pressing button 1 again, the display shows “outP“. (output)
For configuring the outputs of the device, please follow the instructions given in chapter 5.8.