E44.X.00.6C-02 Connection and operating manual GIR 300 page 25 of 28
8 Min-/max-value memory
The device features a minimum/maximum-value storage. In this storage the highest and lowest performance data is
saved.
Calling of the minimum value:
Press button 3 shortly: the device will display “Lo“ briefly, after that the min-value is displayed for about 2 sec.
Calling of the maximum value:
Press button 2 shortly: the device will display “Hi“ briefly, after that the max-value is displayed for about 2 sec.
Erasing of the min/max values:
Press button 2 and 3 for 2 sec.: The device will display “CLr“ briefly, after that the min/max-values are set to the cur-
rent displayed value.
9 Serial interface
The device features one EASYBus-Interface. You can use the device as a full function EASYBus-device.
The serial interface allows the device to communicate with a host computer. Data polling and data transfer is done in
master/slave mode, so the device will only send data on demand. Every device has a unique ID-number that makes
exact identification of each device possible.
With the help of a software (like EASYBus-Configurator freeware version available via internet) you are able to
reassign an address to the device.
Additional accessories needed for the interface mode:
- Interface Converter EASYBUS PC: e.g. EBW 1, EBW 3, EBW 64, EBW 240 and EB 2000 MC
- Software for communication with the device
EBS 20M / 60M: 20- or 60-channel-software for displaying a measured value
EASYControl net: multi-channel software for real-time-recording and displaying measure-values.
EASYBus-DLL: EASYBus-developer-package for developing own software. This package features a universal
WINDOWS®-Library with documentation and program-examples. The DLL can be used in any
usual programming language.
10 Error codes
When detecting an operating state which is not permissible, the device will display an error code.
The following error codes are defined:
Err.1
Exceeding of measuring range
Indicates that the valid measuring range of the device has been exceeded.
Possible causes: - Input signal to high.
- Sensor broken (Pt 100 and Pt 1000).
- Sensor shorted (0(4)-20 mA).
- Counter overflow
Remedies: - The error-message will be reset if the input signal is within the limits.
- check sensor, transducer or transmitter.
- check device configuration (e.g. input signal)
- reset the counter.
Err.2
Values below measuring range
Indicates that the values are below the valid measuring range of the device.
Possible causes: - Input signal is to low or negative.
- Current below 4mA.
- Sensor shorted (Pt 100 and Pt 1000).
- Sensor broken (4-20 mA).
- Counter underflow
Remedies: - The error-message will be reset if the input signal is within the limits.
- Check sensor, transducer or transmitter.
- check device configuration (e.g. input signal)
- Reset the counter.