E44.X.00.6C-02 Connection and operating manual GIR 300 page 26 of 28
Err.3
Display range has been exceeded
Indicates that the valid display range (9999 digit) of the device has been exceeded.
Possible causes: - Incorrect scale.
- Counter overflow
Remedies: - The error-message will be reset if the display value is below 9999.
- Reset the counter.
- When happening frequently, check the scale-setting, maybe it was set
too high and should be reduced
Err.4
Values below display range
Indicates that display value is below the valid display range of the device (-1999 digit).
Possible causes: - Incorrect scale.
- Counter underflow
Remedies: - The error-message will be reset if the display value is above -1999.
- Reset the counter.
- When happening frequently, check the scale-setting, maybe it was set
too low and should be increased
Err.7
System error
The device features an integrated self-diagnostic-function which checks essential parts of the device
permanently. When detecting a failure, error-message Err.7 will be displayed.
Possible causes: - Valid operating temperature range has been exceeded or is below
the valid temperature range.
- Device defective
Remedies: - Stay within valid temperature range
- Exchange the defective device
Er.9
System error
The device features an integrated diagnostic-function for the connected sensor or transmitter.
When detecting a failure, error-message Err.9 will be displayed.
Possible causes: - Sensor broken or sensor shorted (Pt 100 or Pt 1000).
- Sensor broken (thermo-elements)
Remedies: - Check sensor or exchange defective sensor
Er.11
Value could not be calculated
Indicates a measuring value, needed for calculation of the display value, is faulty or out of range.
Possible causes: - Incorrect scale
Remedies: - Check settings and input signal