11 Specification
Absolute maximum ratings: see chapter 4.2
Measuring inputs: Standard inputs for
Input type
Input signal
Measuring range
Resolution
Note
Standard
voltage-signal
0 – 10 V
0 ... 10 V
Ri > 300 kOhm
0 – 2 V
0 ... 2 V
Ri > 10 kOhm
0 – 1 V
0 ... 1 V
Ri > 10 kOhm
0 – 50 mV
0 ... 50 mV
Ri > 10 kOhm
Standard
current-Signal
4 – 20 mA
4 ... 20 mA
Ri = ~ 125 Ohm
0 – 20 mA
0 ... 20 mA
Ri = ~ 125 Ohm
RTD probes
Pt100 (0.1°C)
-50.0 ... +200.0 °C
(or -58.0 ... +392.0 °F)
0.1 °C / °F
3-wire-connection
max. perm. line resistance: 20 Ohm
Pt100 (1°C)
-200 ... +850 °C
(or -328 ... +1562 °F)
1 °C / °F
3-wire-connection
max. perm. line resistance: 20 Ohm
Pt1000 (0.1°C)
-100.0 ... +200.0 °C
(or -148.0 ... +392.0 °F)
0.1 °C / °F
2-wire-connection
Pt1000 (1°C)
-200 ... +850 °C
(or -328 ... +1562 °F)
1 °C / °F
2-wire-connection
Thermocouple
probes
NiCr-Ni (type K)
-270 ... +1350 °C
(or -454 ... +2462 °F)
1 °C / °F
Pt10Rh-Pt (type S)
-50 ... +1750 °C
(or -58 ... +3182 °F)
1 °C / °F
NiCrSi-NiSi (type N)
-270 ... +1300 °C
(or -454 ... +2372 °F)
1 °C / °F
Fe-CuNi (type J)
-170 ... +950 °C
(or -274 ... +1742 °F)
1 °C / °F
Cu-CuNi (type T)
-270 ... +400 °C
(or -454 ... +752 °F)
1 °C / °F
Frequency
TTL-Signal
0 Hz ... 10 kHz
0.001 Hz
signal low: 0.0 – 0.5 V
signal high: 2.7 – 24 V
Switching contact
NPN
0 Hz ... 3 kHz
0.001 Hz
an internal pull-up-resistor (~7 kOhm to
+3.3V) is connected automatically.
Switching contact
PNP
0 Hz ... 1 kHz
0.001 Hz
an internal pull-down-resistor (~7 kOhm
to +3.3V) is connected automatically.
Rotation
TTL-Signal,
Swit.contact NPN, PNP
0 ... 9999 U/min
0.001 U/min
Pre-scaling-factor (1-1000),
Pulse frequency:
max. 600000 p./min. *
Up/Downwards-
Counter
TTL-Signal,
Swit.contact NPN, PNP
0 ... 9999
with pre-scaling factor: 9 999 000
Pre-scaling-factor (1-1000),
Pulse frequency:
max. 10000 p./sec. *
* = with switching contact accordingly to frequency input lower values may occur
Display range: (voltage-, current and frequency-measurement)
-1999 ... 9999 digit, initial value, terminal value and decimal point position arbitrary
Recommended rang: < 2000 digit
Accuracy: (at nominal temperature)
Standard signal: < 0.2% FS ±1 digit (from 0 – 50 mV: < 0.3% FS ±1 digit)
RTD: < 0.5% FS ±1 digit
Thermocouples: < 0.3% FS ±1 digit (from type S: < 0.5% FS ±1 digit)
Frequency: < 0.2% FS ±1 digit
Point of compensation
: ±1°C ±1 digit (at nominal temperature)
Temperature drift: < 0.01% FS / K (from Pt 100 - 0.1°C: < 0.015% FS / K)