Operating Manual
EASYBUS temperature sensor module
EBT – IF... as of V1.4
Specification:
Measuring range: please refer to type plate
EBT – IF1 (standard): -30,0 ... +100,0 °C
EBT – IF2 (standard): -30,0 ... +100,0 °C
EBT – IF3 (standard): -70,0 ... +400,0 °C
Measuring probe: internal Pt1000-sensor
Accuracy:
(at nominal temperature) ±0,2% of meas. value ±0,2°C (EBT-IF1, EBT-IF2)
±0,3% of meas. value ±0,2°C (EBT-IF3)
Min-/Max-value memory: min- and max measured value are stored
Output signal: EASY
BUS-protocol
Connection: 2-wire EASY
BUS, polarity free
Busload: 1.5 EASY
BUS-devices
Adjusting: via interface by input of offset and scale value
Ambient conditions for electronic
(in sleeve):
Nominal temperature: 25°C
Operating temperature: -25 to 70°C
During operation please take care, that even at higher temperatures at the sensor tube ( >70°C)
the allowable temperature range of the electronics, placed in the sleeve, may not exceed!
Relative humidity: 0 to 100 %RH
Storage temperature: -25 to 70°C
Housing: stainless steel housing
Dimensions: depending on sensor construction
Sleeve: Ø 15 x 35 mm (without screwing)
Tube length FL: 100 mm or 50 mm or on customer requirement
Tube diameter D: Ø 6 mm or on customer requirement
(available Ø: 4, 5, 6 and 8 mm)
Collar tube length HL: 100 mm or on customer requirement
Thread: G1/2“ or on customer requirement
(available threads M8x1, M10x1, M14x1.5, G1/8“, G1/4“, G3/8“, G3/4“)
IP rating: IP67
Electrical connection: polarity free connection via 2-pol connection cable
Cable length: 1m or on customer requirement
EMC: The
device corresponds to the essential protection ratings established in the Regulations of the Council for the
Approximation of Legislation for the member countries regarding electromagnetic compatibility (2004/108/EG).
In accordance with
EN61326 +A1 +A2 (appendix A, class B), additional errors: < 1% FS.
The tube have to be protected sufficiently against ESD pulses, if the device is used in areas with risk of ESD.
When connecting long leads adequate measures against voltage surges have to be taken.
Disposal instructions:
The device must not be disposed in the regular domestic waste.
Send the device directly to us (sufficiently stamped), if it should be disposed. We will dispose the device appropriate and
environmentally sound.