4 Measurement Basics
The resistance of the medium between the conical metal
surfaces (1) and (2) is measured. The medium being
measured has to be compressed well enough.
The best is, having a constant pressure onto the medium
during the measurement. Do not release handle during the
measuring, otherwise the contact to the medium can be
interrupted, in this case a to dry value would be
measured.
Attention: Especially at bulk material stored outside or very wet
material, large distribution of the actual moisture values can
appear.
It is best to do several measurements and taking the
average in order to get meaningful results:
The measurement of wood chips or things like that is
depending on temperature. For an exact measuring result the
temperature is automatically compensated when using the
suitable instrument (e.g. GMH 3830). The temperature-
measuring is done at the tip of the probe (1), a sufficient time
to adjust the sensors temperature to the material has to be
waited for.
Different measuring results are depending on different types of
material. Select correct material-group or material type before
measuring. Refer to operation manual of the connected
measuring device. When pushing in the probes, oscillating
movements have to be avoided. Otherwise hollows between
the probes and the material may falsify the measurement.
5 Unit conversion: moisture dry-and wetbasis
Older instruments (e.g. GMH 3830 before V1.4) cannot be switched from moisture content u to w, in this
case the wet basis moisture content can be calculated like following:
Wet-basis moisture w [%] = 100 * Moisture content u [%] / (100 + Moisture content u [%])
Example: 1 kg of wet wood, which contains 500 g of water has a moisture content u of 50%
0
10
20
30
40
50
60
0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150
wet-basis moisture content w [%]
moisture content u [%] = display of GMH 38x0
Conversion moisture content u - wet-basis moisture content w