Operator's Manual

ETD-100 OPERATOR’S MANUAL
19
Figure 11: Insert Swab Screen
6.2 Device User Interface
The device user interface has been designed to be simple and effective for Users. All navigation between menus and
screens on the device are performed using the touchscreen.
The following sections describe the device User Interface.
6.2.1 The INSERT SWAB Screen
The INSERT SWAB Screen is the default screen for the device
when waiting for a test sample.
The horizontal bars on the left of the screen indicate the status
of the: filter, consumable reagent fluids and battery.
Consumable items need to be replaced periodically when
expired or, in the case of the filter, if the baseline of the
electropherogram is ‘noisy’ and the battery needs to be kept
charged. If any indicators drop below a threshold, a prompt will
be displayed indicating that attention is required and the bar will turn red. Refer to Section 8.1 Maintenance Overview.
The SETTINGS BUTTON on the INSERT SWAB Screen is used to access the Settings Screen.
The Settings Screen is described in Section 6.2.2
.
Figure 12: INSERT SWAB Screen (No Labels)