Operator's Manual

ETD-100 OPERATOR’S MANUAL
23
5. After the sample has been tested a results screen will be displayed indicating a Pass (green indicator), a
Warning (red indicator), or an Invalid Sample - Overload (white indicator) as illustrated in Figure 19. The
Swab should then be removed and disposed.
Figure 19: PASS, WARNING and OVERLOAD Screens (Wait)
6. Once the User is ready to move onto the next test, the PROCEED button can be pressed to prepare the
device for the next test. The INSERT SWAB Screen will be displayed.
6.3.2 Results Screens
Following a sample test, the User will be presented with either a Pass Screen, Warning Screen or Invalid Sample
Overload Screen.
If the sample does not contain inorganic material, the PASS Screen will be displayed. The device will automatically
initiate a flush sequence to prepare for a new sample. The User will be able to CONTINUE to the next sample test.
If the sample indicates a sample test alert i.e. one or more of analytes detected: Chlorates, Perchlorates or Nitrates
above a defined level have been found, the WARNING Screen will be displayed. The User will be prompted to
remove the Swab. Once the Swab has been removed and the User has pressed CONTINUE, the device will run an
extended flush followed by a sample test to ensure that the next sample will not be contaminated. If the sample
detects an overload of a substance and does not identify Chlorates, Perchlorates or Nitrates above the defined
threshold, the OVERLOAD screen will be displayed. The User will be prompted to remove the Swab. Once the Swab
NOTE: INITIALLY, THE RESULTS SCREEN WILL PROMPT THE USER TO WAIT.
THIS INDICATES THE SYSTEM IS RUNNING A FLUSH SEQUENCE. USERS MUST WAIT
UNTIL THEY SEE THE PROCEED BUTTON ON THE RESULTS SCREEN TO CONTINUE.