Code
Detail
Remedy
C12
Unexpected low pressure
Call service technician
C13
Unexpected low pressure
Call service technician
C14
Air detected
Replace filter and fluid cartridge
C15
Contamination detected
Call service technician
C16
HV high current
Restart diagnostics
C17
Capillary resistance
Restart diagnostics
C18
HV too high
Call service technician
C19
HV too low
Call service technician
C20
Hardware communication fault
Call service technician
C21
Hardware communication fault
Call service technician
C22
Hardware communication fault
Call service technician
C23
LV low current
Call service technician
C24
Unexpected low pressure
Call service technician
C25
Hardware communication fault
Call service technician
C26
Hardware communication fault
Restart Device
C27
Unexpected error
Call service technician
C28
Hardware communication fault
Restart Device
C29
Hardware communication fault
Restart Device
C30
Hardware communication fault
Restart Device
C31
Fluidic fault
Call service technician
C32
Clamp timeout fault
Call service technician
C33
SMC check failed
Call service technician
C34
Unexpected error
Call service technician
C35
Pump motor fault
Call service technician
C36
Unexpected error
Call service technician
C37
Clamp sensor fault
Call service technician
D15
No BGE detected
Replace fluid cartridge
D16
No BGE detected
Replace fluid cartridge
D17
No BGE detected
Replace fluid cartridge
D18
Capillary unprepared
Restart diagnostics
D19
Capillary blocked
Call service technician
D20
Elution unsuccessful
Replace filter and fluid cartridge
D21
Elution unsuccessful
Replace filter and fluid cartridge
D22
Elution unsuccessful
Replace filter and fluid cartridge
D23
No Elution detected
Replace filter and fluid cartridge
D24
No Elution detected
Replace filter and fluid cartridge
D25
Leak detected
Call service technician
D26
Hardware fault
Call service technician
D27
Unexpected high pressure
Replace fluid cartridge
D28
Unexpected high pressure
Replace fluid cartridge
D29
Unexpected high pressure
Replace fluid cartridge
D30
Unexpected high pressure
Replace fluid cartridge
D31
36 kV check failed
Call service technician
D32
36 kV check failed
Call service technician
D33
5 kV check failed
Call service technician