User's Manual

Appendix
26
Appendix
27
Not applicable
Table 4 Guidance and
MANUFACTURER
S
declaration electromagnetic
IMMUNITY
for
ME EQUIPMENT
and
ME SYSTEMS
that are not
LIFE
-
SUPPORTING
Guidance and manufacturers declaration electromagnetic immunity
I
MMUN ITY
test
IEC 60601
TEST
LEVE L
Compliance
level
Electromagneti c envi ronmen t
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
3 V/m
Recommended separation distance
P
V
=d ]
5,3
[
1
P=d 2.333
800 MHz to 2,5 GHz
where P is the maximum output power rating of the
transmitter in watts (W ) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m ).
Field strengths from fixed RF transmitters, as deter-
mined by an electromagnetic site survey,
a
should be
less than the compliance level in each frequency
range.
b
Interference may occur in the vicinity of equipm ent
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) t elephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the device is used exceeds
the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the device.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V
1
] V/m.
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should assure that it is used in such an environment.
Portable and mobile RF
communications equipment should
be used no closer to any part of the
device, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
P=d
80 MHz to 800 MHz
1.167
Table 6 Recommended separation distances between portable and mobile
RF communications equipment and the
ME EQUIPMENT or ME SYSTEM
for
ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and the device
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the device as recommended
below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
m
Rated maximum output power
of transmitter
W
150 kHz to 80 MHz
P
V
=d ]
5,3
[
1
80 MHz to 800 MHz 800 MHz to 2,5 GHz
10,0
1,0
1
01
001
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic prop
agation is affected by absorption and
reflection from structures, objects and people.
0.2330.117
0.7380.369
2.3331.167
7.3783.690
23.3311.67
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
P=d
2.333
P=d
1.167