User's Manual

13
Disposable pulmonary function meter filter
Discarded immediately after use in each patient
Special reminder: Disposable products must be discarded.
The following parts must be cleaned and disinfected once a day, see the following table
The screen is cleaned and
disinfected once a day
The air head of the flow sensor is
cleaned and disinfected once a day
Caution:
Thorough disinfection of all contaminated parts can avoid potential risks of infection.
Therefore, all parts should be disinfected regularly, and disposable products must be
discarded!
7.1 Removal steps of flow sensor head
1) Press the button of the rear shell and pull out the flow sensor head forcefully at the
same time;
2) Hold the sensor head flat, and carefully separate the front shell of the head and the
back shell of the head in the up and down direction;
3) Remove the silicone parts from the cavity;
4) Open the quick release card lock to both sides;
5) Separate the parts of the cavity up and down to prevent the screen and the sealing ring
from falling.
7.2 Cleaning and disinfection of the parts of the flow sensor head
1) Cleaning and disinfection of the parts of the flow sensor head under normal
temperature and pressure:
Pre-rinse with running distilled water immediately after use to remove residual saliva,
dirt and other stains.
Soak all the cleaned sensor head parts in 2% glutaraldehyde disinfectant solution for