Operation Instruction

25
1. Soak the dry soft cloth with 75%alcohol.
2. Wipe all surfaces of the motor component, power adapter,
charging base and other components with a wet soft cloth for at
least 3minutes.
3. Wipe the surface of the component with a dry soft nap-free
cloth.
Note:
a) The cleaning and disinfection must be performed within
10min before use.
b) The disinfectant used must be used immediately, no foaming
is allowed.
c) In addition to 75% alcohol, you can use non-residue
disinfectants such as Oxytech from Germany, but you must
respect the concentration, temperature and time specied by the
disinfectant manufacturer.
d) After cleaning and disinfecting the motor component, you
must install a disposable barrier before use and repeat steps 1,
2 and 3 to clean the disposable barrier (For detailed installation