Operation Instruction

31
• There should be a ushing connector connected to the inner
cavity of the product.
• The cleaning procedure is suitable for the product, and the
irrigating period is su󰀩cient.
It is recommended to use a washer-disinfector in accordance
with EN ISO 15883. For the specic procedure, please refer
to the automated disinfection section in the next section
"Disinfection".
Notes:
a) The cleaning agent does not have to be pure water. It can
be distilled water, deionized water or multi-enzyme. But please
ensure that the selected cleaning agent is compatible with the
product.
b) In washing stage, the water temperature should not exceed
45
. Otherwise the protein will solidify and it would be di󰀩cult
to remove.
c) After cleaning, the chemical residue should be less than
10mg /L.