Data Sheet

GPT-9500 Series User Manual
128
AFTER FAIL
STOP
1. If After Fail Stop is set for an AUTO test, then
the tester will stop the whole AUTO test
immediately when a FAIL judgment of anyone
of the steps occurs. See page 99 for more details.
Normal
AUTO
Display
for
AFTER
FAIL
STOP
AUTO test stops in the
1st step immediately
AUTO
Page
View
Display
for
AFTER
FAIL
STOP
AUTO test stops in the
1st step immediately
2. The FAIL indicator on the front
panel will also be lit. The buzzer
will sound when activated.
3. When FAIL is displayed on-screen,
press the STOP button to return to
the READY status.
Note
When AFTER FAIL STOP occurs, only the STOP