EXTERNAL CONTROL
193
Using the SIGNAL I/O to Start/Stop Tests
To use the SIGNAL I/O port the CONTROL
settings have to be set to SIGNAL IO in the
UTILITY mode.
1. Set the CONTROL option to Signal
in the UTILITY mode.
Page 154
2. Connect the Input/Output signals
to the SIGNAL I/O port.
3. To start the testing, short the STOP
and INPUT_COM line for a
minimum of 1ms to put the tester
into READY status.
4. To start the testing, short the
START and INPUT_COM lines for
a minimum of 1ms.
5. To stop the testing, temporarily
short the STOP and INPUT_COM
line again.
Even if the GPT-9500 is configured to use the
SIGNAL I/O interface, the STOP button on the
front panel can still be used to stop a test.