Data Sheet

REMOTE CONTROL
239
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP1:AC:TIME:FALL 0
Sets the fall time value for ACW of step1 to off.
[:SOURce]:SAFEty:STEP<n>:AC:TIME:FALL?
Returns the fall time value for ACW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return Parameter: <NR3>
Example: SAFE:STEP1:AC:TIME:FALL?
> +0.000000E+00
The fall time value for ACW of step1 set is off.
[:SOURce]:SAFEty:STEP<n>:AC:CHANnel[:HIGH]
Sets the output channel status for ACW of scanning test high
voltage.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: (@(CH))
Example: SAFE:STEP1:AC:CHAN (@(1,3))
Sets the output channel for ACW of step1 to channel 1 and 3
HIGH output
Example: SAFE:STEP1:AC:CHAN (@(0))
Sets the output channel for ACW of step1 to off for all HIGH
output.
[:SOURce]:SAFEty:STEP<n>:AC:CHANnel[:HIGH]?
Returns the output channel status for ACW of scanning test
high voltage.