> 0
The ground mode for ACW of step1 is off.
[:SOURce]:SAFEty:STEP<n>:DC[:LEVel]
Sets the test voltage value for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NR1> (50 ~ 6000)
Example: SAFE:STEP2:DC:LEV 5000
Sets the test voltage value for DCW of step2 to 5000V.
[:SOURce]:SAFEty:STEP<n>:DC[:LEVel]?
Returns the test voltage value for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: <NR3>
Example: SAFE:STEP2:DC:LEV?
> +5.000000E+03
The test voltage value for DCW of step2 is 5000V.
[:SOURce]:SAFEty:STEP<n>:DC:LIMit[:HIGH]
Sets the leakage current high limit for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP2:DC:LIM 0.009
Sets the leakage current high limit for DCW of step2 to 9mA.
[:SOURce]:SAFEty:STEP<n>:DC:LIMit[:HIGH]?
Returns the leakage current high limit for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)