Data Sheet

REMOTE CONTROL
243
Return parameter: <NR3>
Example: SAFE:STEP2:DC:LIM?
> +9.000000E-03
The leakage current high limit for DCW of step2 is 9mA.
[:SOURce]:SAFEty:STEP<n>:DC:LIMit:LOW
Sets the leakage current low limit for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP2:DC:LIM:LOW 0.0001
Sets the leakage current low limit for DCW of step2 to 0.1mA.
[:SOURce]:SAFEty:STEP<n>:DC:LIMit:LOW?
Returns the leakage current low limit for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: <NR3>
Example: SAFE:STEP2:DC:LIM:LOW?
> +1.000000E-04
The leakage current low limit for DCW of step2 is 0.1mA.
[:SOURce]:SAFEty:STEP<n>:DC:LIMit:ARC[:LEVel]
Sets the ARC value for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP2:DC:LIM:ARC 0.006
Sets the ARC value for DCW of step2 to 6mA.