Data Sheet

REMOTE CONTROL
245
Example: SAFE:STEP2:DC:TIME:RAMP 0.3
Sets the ramp time value for DCW of step2 to 0.3s.
[:SOURce]:SAFEty:STEP<n>:DC:TIME:RAMP?
Returns the ramp time value for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: <NR3>
Example: SAFE:STEP2:DC:TIME:RAMP?
> +3.000000E-01
The ramp time value for DCW of step2 set is 0.3s.
[:SOURce]:SAFEty:STEP<n>:DC:TIME[:TEST]
Sets the test time value for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP2:DC:TIME 2
Sets the test time value for DCW of step2 to 2s.
[:SOURce]:SAFEty:STEP<n>:DC:TIME[:TEST]?
Returns the test time value for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: <NR3>
Example: SAFE:STEP2:DC:TIME?
> +2.000000E+00
The test time value for DCW of step2 set is 2s.